Tap for Standard and Tapered Implant

Additional use of drill taps to reduce bone compression in very hard bone

The Aadva Implant Line has a full line up of dental implant drivers and tools, guaranteeing a precise implant placement and tightening of the implant parts.

Standard

Narrow – Short

810372

Narrow – Long

810373

Regular – Short

810374

Regular – Long

810375

Wide – Short

810376

Wide – Long

810380

Tapered

Narrow – Short

810381

Narrow – Long

810382

Regular – Short

810383

Regular – Long

810384

Wide – Short

810385

Wide – Long

810386

Advice for use of Drill Taps

  • Use with sufficient cooling by saline during tapping and at speeds of 25 rpm or less.
  • Use at the end in a drill sequence.
  • Use with attention to tilt.
  • Increase handpiece setting torque gradually (MAX: 50 Ncm)
  • If Tap has not reached the desired depth with 50 Ncm repeat forward/reverse tapping.
  • After desired depth is reached, tap is removed by reverse rotation under saline irrigation.
  • Tap retrieval should be done in a straight line to prevent damage to bone screw channels.

Tap for Standard and Tapered Implant - Position of laser marking

  • For standard implant
  • For tapered implant

* Except for 8 mm line, position of laser marking of Tapered Implant Tap is different from Twist Drill and Tapered Implant Drill in order to enable the appropriate insertion torque of implant. Therefore, except for 8 mm, the tapping depth is shallower than the implant length.